Key result
Implantation of the Carpentier-Edwards Perimount Magna Ease aortic valve bioprosthesis resulted in an absence of prosthesis-patient mismatch in 99.5% of patients.
Why the study?
Does the Carpentier-Edwards Perimount Magna Ease aortic valve bioprosthesis prevent prosthesis-patient mismatch in patients undergoing aortic valve replacement?
Cohort (n=270)
Does the Carpentier-Edwards Perimount Magna Ease aortic valve bioprosthesis prevent prosthesis-patient mismatch in patients undergoing aortic valve replacement?
The Carpentier-Edwards Perimount Magna Ease aortic valve bioprosthesis demonstrates excellent hemodynamic performance with an incidence of severe prosthesis-patient mismatch below 0.5%.
Low mismatch rates support hemodynamic performance in AVR; leaves open comparative outcomes versus alternatives in randomized trials.
Prosthesis-patient mismatch (PPM) remains a controversial issue with the most recent stented biological valves. We analyzed the incidence of PPM after implantation of the Carpentier-Edwards Perimount Magna Ease aortic valve (PMEAV) bioprosthesis and assessed the early clinical outcome. Two hundred and seventy consecutive patients who received a PMEAV bioprosthesis between January 2007 and July 2008 were analyzed. Pre-, peri- and postoperative data were assessed and echocardiographic as well as clinical follow-up was performed. Mean age was 72+/-9 years, 168 (62.2%) were males. Fifty-seven patients (21.1%) were below 65 years of age. Absence of PPM, corresponding to an indexed effective orifice area >0.85 cm(2)/m(2), was 99.5%. Observed in-hospital mortality was 2.2% (six patients), with a predicted mortality according to the additive EuroSCORE of 7.6+/-3.1%. At echocardiographic assessment after a mean follow-up period of 150+/-91 days, mean transvalvular gradient was 11.8+/-4.8 mmHg (all valve sizes). No paravalvular leakage was seen. Nine patients died during follow-up. The Carpentier-Edwards PMEAV bioprosthesis shows excellent hemodynamic performance. This valve can be implanted in all sizes with an incidence of severe PPM below 0.5%.
No takes yet. Share an insight, caveat, or question.
Wyss et al. (2010) conducted a cohort in Aortic valve replacement (n=270). Carpentier-Edwards Perimount Magna Ease aortic valve (PMEAV) bioprosthesis was evaluated on Absence of prosthesis-patient mismatch (indexed effective orifice area >0.85 cm2/m2). Implantation of the Carpentier-Edwards Perimount Magna Ease aortic valve bioprosthesis resulted in an absence of prosthesis-patient mismatch in 99.5% of patients.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: