Key result
Severe prosthesis-patient mismatch after AVR is linked to ~38% higher late overall mortality.
Why the study?
Following findings that prosthesis-patient mismatch predicts operative mortality after AVR, its effect on late survival and whether age, BMI, and preoperative LV function modulate this impact remained to be evaluated.
Does prosthesis-patient mismatch increase late mortality in patients who have survived aortic valve replacement?
Cohort (n=2,576)
Does prosthesis-patient mismatch increase late mortality in patients who have survived aortic valve replacement?
Answers are grounded in this paper and cite their sources.
Severe prosthesis-patient mismatch increases late mortality after aortic valve replacement, particularly in younger, non-obese patients, while moderate mismatch increases mortality in those with left ventricular dysfunction.
OBJECTIVES: This study was designed to evaluate the effect of valve prosthesis-patient mismatch (PPM) on late survival after aortic valve replacement (AVR) and to determine if this effect is modulated by patient age, body mass index (BMI), and pre-operative left ventricular (LV) function. BACKGROUND: We recently reported that PPM is an independent predictor of operative mortality after AVR, particularly when associated with LV dysfunction. METHODS: The indexed valve effective orifice area (EOA) was estimated in 2,576 patients having survived AVR and was used to define PPM as not clinically significant if it was >0.85 cm(2)/m(2), as moderate if >0.65 and < or =0.85 cm(2)/m(2), and severe if < or =0.65 cm(2)/m(2). RESULTS: After adjustment for other risk factors, severe PPM was associated with increased late overall mortality (hazard ratio [HR]: 1.38; p = 0.03) and cardiovascular mortality (HR: 1.63; p = 0.0006) in the whole cohort. Severe PPM was also associated with increased overall mortality in patients <70 years old (HR: 1.77; p = 0.002) and in patients with a BMI <30 kg/m(2) (HR: 2.1; p = 0.006), but had no impact in older patients or in obese patients. Moderate PPM was a predictor of mortality in patients with LV ejection fraction <50% (HR: 1.21; p = 0.01), but not in patients with preserved LV function. CONCLUSIONS: Moderate PPM is associated with increased late mortality in patients with LV dysfunction, but with normal prognosis in those with preserved LV function. Notwithstanding the previously demonstrated deleterious effect of severe PPM on early mortality, this factor appears to increase late mortality only in patients <70 years old and/or with a BMI <30 kg/m(2) or an LV ejection fraction <50%.
No takes yet. Share an insight, caveat, or question.
Mohty et al. (2008) conducted a cohort in Aortic valve replacement (n=2,576). Severe prosthesis-patient mismatch vs. Not clinically significant prosthesis-patient mismatch was evaluated on Late overall mortality (HR 1.38, p=0.03). Severe prosthesis-patient mismatch was associated with increased late overall mortality (HR 1.38; p=0.03) and cardiovascular mortality (HR 1.63; p=0.0006) after aortic valve replacement.
Synapse has enriched 10 related papers. Consider them for comparative context: