Why the study?
How can MRI be safely performed in patients with cardiac implantable electronic devices?
How can MRI be safely performed in patients with cardiac implantable electronic devices?
This document provides endorsed protocols for safely performing MRI scans in patients with cardiac implantable electronic devices, including those with non-MR conditional devices under specific circumstances.
Answers are grounded in this paper and cite their sources.
May expand MR access for select CIED patients; leaves open need for prospective safety data in non-conditional devices.
MR scanning in patients with cardiac implantable electronic devices (CIEDs) was formerly felt to be contraindicated, but an increasing number of patients have an implanted MR conditional device, allowing them to safely undergo MR scanning, provided the manufacturer's guidance is adhered to. In addition, some patients with non-MR conditional devices may undergo MR scanning if no other imaging modality is deemed suitable and there is a clear clinical indication for scanning which outweighs the potential risk. The following guidance has been formulated by the British Heart Rhythm Society and endorsed by the British Cardiovascular Society and others. It describes protocols that should be followed for patients with CIEDs undergoing MR scanning. The recommendations, principles and conclusions are supported by the Royal College of Radiologists.
No takes yet. Share an insight, caveat, or question.
Löwe et al. (2015) studied Cardiac implantable electronic devices (CIEDs). MR scanning was evaluated. MR scanning can be safely performed in patients with cardiac implantable electronic devices when following specific protocols outlined by the British Heart Rhythm Society.
Synapse has enriched 10 related papers. Consider them for comparative context: