Answers are grounded in this paper and cite their sources.
INTRODUCTION: The enumeration and differentiation of nuclear elements in synovial fluid is a cornerstone for diagnosis and follow-up of many orthopedic and rheumatologic diseases. In this study, we evaluated the analytical performance of Mindray BC-6800 BF mode (BC-6800-BF) for synovial fluid analysis. METHODS: Overall, 78 synovial fluids were collected and analyzed with both BC-6800-BF and light microscopy. The study also entailed the assessment of limit of blank (LoB), limit of detection (LoD), limit of quantification (LoQ), carryover and linearity. RESULTS: cells/L, respectively. Linearity was excellent and carryover was negligible. The agreement between BC-6800-BF and light microscopy was satisfactory for all samples pretreated with hyaluronidase, displaying a bias between -5.9% and 8.2%. CONCLUSIONS: The use of BC-6800-BF for synovial fluid analysis enables rapid and accurate assessment, especially for total cell and polymorphonuclear counts. The use of BC-6800-BF may therefore allow the replacement of optical analysis, especially in samples pretreated with hyaluronidase, thus allowing its routine use for the screening of synovial specimens.
No takes yet. Share an insight, caveat, or question.
Buoro et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: