Key result
Self-monitoring with a smartphone ECG and cloud platform did not increase AF detection compared to standard follow-up but reduced the need for additional monitors.
Why the study?
Clinical adoption of direct-to-consumer ECG devices has been limited by the lack of efficient workflow, prompting evaluation of a care model using smartphone ECG coupled with a novel cloud-based platform after AF ablation.
Does self-monitoring using a smartphone ECG coupled with a cloud-based platform improve AF detection and reduce healthcare utilization in patients post-AF ablation compared to standard clinical follow-up?
RCT (n=100)
Does self-monitoring using a smartphone ECG coupled with a cloud-based platform improve AF detection and reduce healthcare utilization in patients post-AF ablation compared to standard clinical follow-up?
Answers are grounded in this paper and cite their sources.
Smartphone ECG monitoring with a cloud-based platform in post-AF ablation patients reduces the need for traditional cardiac monitors without increasing patient anxiety.
Background Direct-to-consumer devices allow patients to record electrocardiograms (ECG) and detect atrial fibrillation (AF). Clinical adoption of these devices has been limited owing to the lack of efficient workflow. Objective To assess a new care model for following patients after AF ablation that uses a smartphone ECG coupled with a novel cloud-based platform. Methods This was a pilot study to describe AF detection, healthcare utilization, use of additional ECGs and cardiac monitors, and changes in anxiety after AF ablation. Patients presenting 3–4 months after early successful AF ablation were randomized into a control group with standard clinical follow-up or a self-monitoring group using smartphone ECG (Kardia Mobile, KM) coupled with a cloud-based platform (KardiaPro, KP) that alerted the physician when AF was detected and followed for 6 months Results A total of 100 patients were randomized: 51 to the KM/KP group and 48 to the control group (1 withdrew). AF was detected in 18 patients (18.2%), 11 (21.6%) in the KM/KP group and 7 (14.6%) in the control group ( P = .42). AF detection occurred at a median of 68 and 91 days in the KM/KP and control groups, respectively ( P = .93). These differences were not statistically significant. Healthcare utilization and changes in anxiety were similar between the groups. More patients required additional ECGs or cardiac monitors in the control group (27.1%) compared to the KM/KP group (5.9%) ( P = .004). Conclusions Smartphone ECG with a cloud-based platform can be incorporated into the care of post–AF ablation patients without increasing anxiety and with less need for additional traditional monitors.
No takes yet. Share an insight, caveat, or question.
Lambert et al. (2021) conducted an RCT in Atrial fibrillation post-ablation (n=100). Smartphone ECG (Kardia Mobile) with cloud-based platform (KardiaPro) vs. Standard clinical follow-up was evaluated on Atrial fibrillation detection (p=0.42). Self-monitoring with a smartphone ECG and cloud platform did not increase AF detection compared to standard follow-up but reduced the need for additional monitors.
Synapse has enriched 10 related papers. Consider them for comparative context: