Key result
Bipolar epicardial wires resulted in no ventricular sensing failures compared to 17.6% with unipolar wires, and significantly improved sensing magnitude (11.13 vs 5.65 mV, P<0.001).
Why the study?
Does a bipolar epicardial wire improve pacing thresholds and sensing compared to a unipolar wire in patients undergoing coronary artery surgery?
Does a bipolar epicardial wire improve pacing thresholds and sensing compared to a unipolar wire in patients undergoing coronary artery surgery?
Bipolar epicardial wires provide more reliable temporary ventricular pacing and sensing post-cardiac surgery compared to unipolar wires.
Answers are grounded in this paper and cite their sources.
Bipolar wires were associated with fewer ventricular sensing failures after coronary surgery; leaves open whether they should supplant unipolar leads pending randomized trials.
The study compared the clinical reliability of using a bipolar epicardial wire (6495, Medtronic) over a unipolar type (FEP15, Ethicon) for post-operative pacing in coronary artery surgery. Atrial and ventricular wires of both types were implanted in 18 patients. Sensitivities and pacing thresholds were tested for 5 consecutive days. Results show that pacing thresholds were better maintained with the bipolar wire in both atria and ventricles. However, sensing failures were frequent in the atrial position (34% vs 9.3% compared with unipolar). By contrast, in the ventricle, no sensing failures occurred (0% vs 17.6% compared with unipolar). Furthermore, sensing magnitude was significantly better (11.13+/-1.32 vs 5.65+/-0.53 mV, P<0.001). We conclude that a single 6495 bipolar wire is effective for temporary ventricular pacing, whilst double unipolar wires remain a useful strategy for securing atrial sensing and pacing.
No takes yet. Share an insight, caveat, or question.
Patrick Yiu (2001) studied coronary artery surgery (n=18). Bipolar epicardial wire (6495, Medtronic) vs. Unipolar epicardial wire (FEP15, Ethicon) was evaluated on Ventricular sensing failures. Bipolar epicardial wires resulted in no ventricular sensing failures compared to 17.6% with unipolar wires, and significantly improved sensing magnitude (11.13 vs 5.65 mV, P<0.001).