Key result
TAV crimping stress decreases leaflet elastic modulus at 2 MPa, establishing a tissue damage threshold.
Why the study?
Leaflet damage occurs during transcatheter aortic valve deployment due to mechanical crimping loads, but the mechanisms of traumatic leaflet tissue damage remained unclear.
Compressive stress exceeding 2 MPa during TAV crimping causes significant structural damage to bovine pericardium leaflets, highlighting the need to optimize crimping size.
Crimping forces >2 MPa may damage TAV leaflets; hypothesis-generating for optimized protocols before clinical adoption.
Leaflet damage has been documented to occur while deploying a transcatheter aortic valve (TAV) due to mechanical loads during the crimping procedures. In this study, the impact of compressive stress on folded leaflets was measured to investigate the mechanism of traumatic leaflet tissue damage. Numerical simulation of TAV crimping procedure was adapted to calculate stress magnitude and distribution of leaflets. A 20 mm balloon expanding short stent TAV with 0.25 mm thickness leaflets was used in the simulation. Then the calculated stresses were applied on leaflet material (bovine pericardium) samples by loading experiments. Mechanical properties evaluation combined with histological and microscopy observation were used to investigate the tissue damage. The elastic modulus and the tensile strength of the tissue began to decrease significantly at 2 MPa stress and 2.5 MPa stress, respectively. No significant differences were observed at 0-1.5 MPa stress. When the TAV was crimped to 14 Fr and 12 Fr, the 2 MPa greater areas on leaflets increased from 18.17% to 76.96%. 2 MPa compressive stress might be the threshold value for leaflet damage. The TAV crimping size should be paid attention to avoid the compressive stress higher than 2 MPa.
No takes yet. Share an insight, caveat, or question.
Feng et al. (2022) studied Transcatheter aortic valve leaflet damage. Compressive stress during TAV crimping was evaluated on Tissue damage (elastic modulus and tensile strength). Compressive stress during transcatheter aortic valve crimping significantly decreased leaflet elastic modulus at 2 MPa and tensile strength at 2.5 MPa, establishing a threshold for tissue damage.