Key result
Precordial magnet and VOO reprogramming fail to prevent pacemaker inhibition and asystole during unipolar electrosurgery.
Why the study?
Does a precordial magnet or VOO reprogramming prevent pacemaker inhibition from unipolar electrosurgical units in pacemaker-dependent patients?
Population
1 pacemaker-dependent 70-year-old man with a DDDR pacemaker placed for complete heart block, undergoing neck…
Design
Case_report
Authors
Loading...
Standard magnet or VOO reprogramming may fail to prevent asystole from unipolar electrosurgery in select devices; hypothesis-generating for device-specific protection circuits.
Case Report (n=1)
No
Does a precordial magnet or VOO reprogramming prevent pacemaker inhibition from unipolar electrosurgical units in pacemaker-dependent patients?
In some pacemakers with high-voltage protection circuits, unipolar ESU can cause temporary pacing failure and asystole that is not prevented by a precordial magnet or VOO reprogramming, necessitating the use of a bipolar ESU or alternative scalpel.
Kleinman et al. (1997) conducted a case report in Pacemaker dependence (n=1). Precordial magnet and pacemaker programmer was evaluated on Pacemaker inhibition and asystole. Electromagnetic interference from a unipolar electrosurgical unit caused total pacemaker inhibition and asystole despite proper placement of a precordial magnet and reprogramming to VOO mode.