Key result
Coagulation-mode surgical electrocautery linked to intraoperative electromagnetic interference and asystole in a DDDR pacemaker patient.
Why the study?
Electromagnetic interference in pacemakers has almost always been reported with the cutting mode of monopolar electrocautery and rarely with the coagulation mode, creating uncertainty about risks during coagulation.
Does surgical electrocautery in the coagulation mode cause electromagnetic interference in patients with pacemakers?
Case Report (n=1)
Does surgical electrocautery in the coagulation mode cause electromagnetic interference in patients with pacemakers?
Clinicians should be aware that the coagulation mode of electrocautery can cause electromagnetic interference and hemodynamic instability in patients with pacemakers.
Answers are grounded in this paper and cite their sources.
Supports awareness of coagulation-mode electrocautery risks in pacemaker patients; leaves open need for prospective incidence data.
Electromagnetic interference in pacemakers has almost always been reported in association with the cutting mode of monopolar electrocautery and rarely in association with the coagulation mode. We report a case of electrocautery-induced electromagnetic interference with a DDDR pacemaker (dual-chamber paced, dual-chamber sensed, dual response to sensing, and rate modulated) in the coagulating and not cutting mode during a spine procedure. We also discuss the factors affecting intraoperative electromagnetic interference. A 74-year-old man experienced intraoperative electromagnetic interference that resulted in asystole caused by surgical electrocautery in the coagulation mode while the electrodispersive pad was placed at different locations and distances from the operating site (This electromagnetic interference did not occur during the use of the cutting mode). However, because of careful management, the outcome was favorable. Clinicians should be aware that the coagulation mode of electrocautery can cause electromagnetic interference and hemodynamic instability. Heightened vigilance and preparedness can ensure a favorable outcome.
No takes yet. Share an insight, caveat, or question.
Abdelmalak et al. (2011) conducted a case report in Electromagnetic interference in a cardiac pacemaker (n=1). Surgical electrocautery in coagulation mode vs. Surgical electrocautery in cutting mode was evaluated on Electromagnetic interference resulting in asystole. Surgical electrocautery in the coagulation mode caused intraoperative electromagnetic interference resulting in asystole in a patient with a DDDR pacemaker.