Key result
Middle-amplitude PPG fiducial points optimize PRV accuracy with RMSSD errors under ~3% supine.
Why the study?
Pulse wave detection accuracy depends on PPG morphology, creating a need to define the most accurate fiducial point for pulse rate variability analysis as a surrogate for heart rate variability under non-stationary conditions across sensor locations.
Does selection of specific fiducial points (such as the middle-amplitude point) from forehead and finger photoplethysmographic signals improve the accuracy and reliability of pulse rate variability analysis as a surrogate for electrocardiogram-derived heart rate variability in healthy young adults undergoing a tilt-table test?
Observational (n=18)
No
Does selection of specific fiducial points (such as the middle-amplitude point) from forehead and finger photoplethysmographic signals improve the accuracy and reliability of pulse rate variability analysis as a surrogate for electrocardiogram-derived heart rate variability in healthy young adults undergoing a tilt-table test?
Answers are grounded in this paper and cite their sources.
For accurate pulse rate variability analysis as a surrogate for ECG-derived heart rate variability under non-stationary conditions, the middle-amplitude fiducial point (nM) from infrared forehead PPG signals provides the highest reliability and lowest error.
Objective : The aim of this work is to evaluate and compare five fiducial points for the temporal location of each pulse wave from forehead and finger photoplethysmographic (PPG) pulse wave signals to perform pulse rate variability (PRV) analysis as a surrogate for heart rate variability (HRV) analysis. Approach : Forehead and finger PPG signals were recorded during a tilt-table test simultaneously with the electrocardiogram (ECG). Artefacts were detected and removed and five fiducial points were computed: apex, middle-amplitude and foot points of the PPG signal, apex point of the first derivative signal and the intersection point of the tangent to the PPG waveform at the apex of the derivative PPG signal and the tangent to the foot of the PPG pulse, defined as the intersecting tangents method. Pulse period (PP) time interval series were obtained from both PPG signals and compared with the RR intervals obtained from the ECG. HRV and PRV signals were estimated and classical time and frequency domain indices were computed. Main results : The middle-amplitude point of the PPG signal ( n M ), the apex point of the first derivative ( ), and the tangent intersection point ( n T ) are the most suitable fiducial points for PRV analysis, resulting in the lowest relative errors estimated between PRV and HRV indices and higher correlation coefficients and reliability indices. Statistically significant differences according to the Wilcoxon test between PRV and HRV signals were found for the apex and foot fiducial points of the PPG, as well as the lowest agreement between RR and PP series according to Bland–Altman analysis. Hence, these signals have been considered less accurate for variability analysis. In addition, the relative errors are significantly lower for n M and using Friedman statistics with a Bonferroni multiple-comparison test, and we propose that n M is the most accurate fiducial point. Based on our results, forehead PPG seems to provide more reliable information for a PRV assessment than finger PPG. Significance : The accuracy of the pulse wave detection depends on the morphology of the PPG. There is therefore a need to widely define the most accurate fiducial point for performing a PRV analysis under non-stationary conditions based on different PPG sensor locations and signal acquisition techniques.
No takes yet. Share an insight, caveat, or question.
Peralta et al. (2019) conducted an observational in Healthy volunteers undergoing autonomic nervous system assessment via tilt-table test (n=18). Pulse rate variability (PRV) analysis from forehead and finger photoplethysmography (PPG) using five fiducial points vs. Heart rate variability (HRV) from simultaneous 3-lead electrocardiogram (ECG) was evaluated on Relative errors, correlation, and reliability between PRV and ECG-derived HRV time and frequency domain indices across five PPG fiducial points (p=<0.01). Middle-amplitude PPG fiducial point (nM) yielded the lowest PRV-to-HRV relative errors and high correlation.