Key result
3D-printed simulation guides successful aortic arch repair in a coarctation patient by predicting absent infolding.
Case Report (n=1)
Preoperative 3D printed model simulation can be a useful tool for planning complicated stent grafting in complex aortic anatomy.
Answers are grounded in this paper and cite their sources.
May aid planning in select complex aortic cases; leaves open prospective validation of 3D simulation.
Stent grafting for complex aortic anatomy remains a challenge. In particular, stent graft collapse (i.e. infolding) is possible when an excessive oversized device is needed. We describe a case of preoperative stent grafting simulation using a three-dimensional (3D) printed model for extensive aortic arch repair in a 69-year-old woman with multiple aneurysms combined with coarctation. The patient was scheduled to undergo staged hybrid repair. A stent graft larger than 28 mm in diameter was needed to deploy into a coarctation of 15 mm in diameter during the 2nd stage of the operation. Preoperative, experimental stent grafting using a 3D printed model indicated that infolding would likely not occur. Therefore, we proceeded with surgery, which was successful. This technology could be a useful application for planning complicated stent grafting.
No takes yet. Share an insight, caveat, or question.
Shijo et al. (2018) conducted a case report in Multiple aneurysms combined with coarctation (n=1). Preoperative stent grafting simulation using a 3D printed model was evaluated on Stent graft collapse (infolding) and surgical success. Preoperative stent grafting simulation using a 3D printed model successfully predicted the absence of infolding, allowing for successful extensive aortic arch repair in a patient with coarctation.
Synapse has enriched 10 related papers. Consider them for comparative context: