The performance of multi pixel photon counters (MPPC) of 3 mm × 3 mm size, with 14400 and 3600 pixels, were studied by means of the signal from a laser light pulser and using the 3 mm × 3 mm × 20 mm LSO pixel scintillator. Special attention was paid to measure number of fired pixels, generated by the light of pulser and that of the LSO crystal, using a direct method of a comparison of the light peak position in the pulse height spectrum with that of the single photoelectron. The tests of the LSO crystal showed 1550 ± 80 fired pixels per MeV in the MPPC with 14400 pixels assuring a good linearity of the response up to about 1 MeV energy of gamma rays absorbed in the LSO crystal. Energy resolution of 14.8% for 662 keV gamma rays from137Cs source and a time resolution of about 850 ps for 511 keV annihilation quanta were limited by a rather low number of the fired pixels compared to the number of photoelectrons in photomultipliers.
No takes yet. Share an insight, caveat, or question.
Nassalski et al. (2010) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: