Key result
Biventricular pacemaker implantation was successful in 91% of patients, with satisfactory and stable left ventricular lead pacing thresholds up to 3 months.
Why the study?
Is biventricular pacemaker implantation feasible and safe in a new pacing programme?
Cohort (n=54)
Is biventricular pacemaker implantation feasible and safe in a new pacing programme?
Answers are grounded in this paper and cite their sources.
Biventricular pacemaker implantation is feasible and safe with good medium-term results when instituted by experienced operators.
AIMS: To describe the institution of a biventricular pacing programme with particular reference to the implantation procedure and new technological considerations. METHODS AND RESULTS: Analysis of outcomes of a consecutive series of 54 patients undergoing attempted biventricular pacemaker implantation between February 1998 and April 1999. Successful implantation was achieved in 49 of the 54 patients (91%). Five patients required lead repositioning after initial successful implantation. Left ventricular lead pacing thresholds were found to be satisfactory and stable in the long term, with pacing thresholds of 1.3 V, 1.9 V and 1.6 V at implantation, 1 and 3 months, respectively. Left-sided lead function was not dependent on lead position within the tributaries of the coronary sinus. The implantation procedure was found to be safe, although one patient died during long-term follow-up. CONCLUSION: With appropriate previous experience in complex and coronary sinus pacing and with access to up-to-date pacemaker and lead technology, a biventricular pacemaker implantation service can be instituted with good medium-term results.
No takes yet. Share an insight, caveat, or question.
S. Walker (2000) conducted a cohort in Patients requiring biventricular pacemaker (n=54). Biventricular pacemaker implantation was evaluated on Successful implantation. Biventricular pacemaker implantation was successful in 91% of patients, with satisfactory and stable left ventricular lead pacing thresholds up to 3 months.
Synapse has enriched 10 related papers. Consider them for comparative context: