Key result
Unintentional advancement of an implantable cardioverter defibrillator lead into the left heart caused a transient ischaemic attack in a 70-year-old man.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Unintentional placement of pacemaker or ICD leads in the left heart is an underdiagnosed complication that can present as a transient ischaemic attack.
A frequently underdiagnosed complication of pacemaker and implantable cardioverter defibrillator lead implantation is the unintentional advancement of the leads into the systemic circulation. We report a case encountered in our clinic in a 70-year old man evaluated in a neighbouring clinic with symptoms of transient ischaemic attack with initially unclear aetiology. Posterior-anterior chest X-rays suggested that the lead was in the left heart. This finding was confirmed by transthoracic and transoesophageal echocardiography.
No takes yet. Share an insight, caveat, or question.
Alozie et al. (2011) conducted a case report in Transient ischaemic attack (n=1). Implantable cardioverter defibrillator lead in the left heart was evaluated on Transient ischaemic attack. Unintentional advancement of an implantable cardioverter defibrillator lead into the left heart caused a transient ischaemic attack in a 70-year-old man.