A new rotating voltammetric membrane electrode has been developed and used to measure chlorine dioxide in water and wastewater. It is capable of detecting low concentrations (<1 mg/L) of chlorine dioxide and is suitable for continuous, automated monitoring. At a chlorine dioxide concentration of 0.30 mg/L, the relative standard deviation is 6.4 percent. This electrode has been compared with the chlorophenol red and iodometric methods for measuring chlorine dioxide.
No takes yet. Share an insight, caveat, or question.
Smart et al. (1982) studied this question.