Key result
J-shaped atrial leads show similar overall lead-related complications to straight leads despite differing specific types.
Why the study?
Does a J-shaped atrial lead improve long-term performance and reduce complications compared to a straight atrial lead in patients requiring atrial pacing?
RCT (n=208)
Does a J-shaped atrial lead improve long-term performance and reduce complications compared to a straight atrial lead in patients requiring atrial pacing?
Answers are grounded in this paper and cite their sources.
J-shaped and straight active-fixation atrial leads have comparable overall complication rates at 5 years, but straight leads are more prone to dislodgment while J-shaped leads have more electrical malfunctions.
OBJECTIVE: In this prospective, randomized, control study we compared performance of J-shaped (JL) and straight atrial leads (SL) with active (screw-in) fixation mechanism, over a 5-year follow-up period. SUBJECTS AND METHODS: A total of 208 consecutive patients were randomized for implantation with a Medtronic 4568 JL model (n = 103) versus a Medtronic 4068 SL model (n = 105), and all lead-related measurements and complications were recorded. RESULTS: Lead-related complications occurred in JL and SL with similar frequency (6% and 9%, respectively, P = 0.45). Lead dislodgment occurred in 8 (7.8%) cases in SL versus none in JL (P = 0.004). Lead malfunction and excessive pacing thresholds without macrodislodgment occurred more frequently in JL (11; 10.7%) than in SL (4; 3.8%), P = 0.055. Lead material failure (insulation break) occurred only once in SL at the 5-year checkup. Most complications occurred during the first year after implantation. Other events occurred in both leads with similar frequency and included: death in 39 and 35, loss of follow-up in 8 and 6, and development of atrial fibrillation in 9 and 15 patients, respectively. Of those who completed follow-up, lead impedance and pacing threshold increased significantly in both groups. P wave decreased in SL but not in JL. CONCLUSION: More dislocations occurred in SL, but were counterbalanced by more lead malfunctions and excessive pacing thresholds in the JL group. Both leads demonstrated minor deterioration in electrical measurements during follow-up, with only 1 case of lead material failure in SL. Overall, both groups demonstrated favorable lead performance throughout follow-up.
No takes yet. Share an insight, caveat, or question.
Luria et al. (2005) conducted an RCT in Patients requiring atrial pacing leads (n=208). J-shaped atrial leads (Medtronic 4568 JL) vs. Straight atrial leads (Medtronic 4068 SL) was evaluated on Lead-related complications (p=0.45). J-shaped atrial leads resulted in similar overall lead-related complications compared to straight leads over 5 years, though specific complication types differed.