Why the study?
Does a carbon-coated atrial screw-in lead provide reliable long-term electrical performance and low complication rates in patients requiring atrial pacing?
Does a carbon-coated atrial screw-in lead provide reliable long-term electrical performance and low complication rates in patients requiring atrial pacing?
Carbon-coated atrial screw-in leads demonstrate reliable long-term electrical performance and a low complication rate compared to general atrial leads.
Answers are grounded in this paper and cite their sources.
May support half-nominal output programming in selected patients; leaves open need for randomized confirmation of long-term reliability.
The aim of this study was to investigate long‐term performance of a carbon coated atrial screw‐in lead. During implantation of 247 leads of this type we measured an average stimulation threshold of 0.74 V(range: 0.2–1.6 V) at 1 ms pulse width. Mean lead impedance came to 446 ohms (range: 263–1000 ohms) resulting in an arithmetical energy consumption of 1.51 μJ(range: 0.1–7.21 μJ). Average P wave amplitude was 3.9 mV (range: 1.3–11 mV). After a mean follow‐up of 16.4 months (range; 3–60 months) we found excellent threshold results in 76% of the patients permitting a safety programming al half of nominal value. An additional 14% nominal settings could be retained. With regard to chronic lead impedance of 488 ohms (range: 315–1327 ohms) we calculated an average chronic energy consumption of 10.83 μJ (range: 1.62–22.78 μJ) during safety programming. This made up 34.6% of the corresponding energy consumption during nominal programming. Eighty percent of the patients showed chronic P waves above 2 mV; nearly half of them (n ‐ 94 = 38%) showed a proper sensing function even when programmed to minimal sensitivity settings or above 4 mV. In 19 leads (8%) we observed unsatisfying threshold results requiring high output programmings. All threshold increases occurred within the first year, 84% (n = 16/19) within the first 3 postoperative months. An additional five leads (2%) were found to have a loss of capture, and one (0.5%) a loss of sensing. These complications required reoperation as well as one dislocation (0.5%) resulting in a low complication rate of 3% as compared to the total complication rate of all implanted atrial leads which was substantially higher (n = 48/652 = 7.5%). With regard to our results we consider the carbon coated atrial screw‐in lead to have proven long‐term reliability.
No takes yet. Share an insight, caveat, or question.
Markewitz et al. (1988) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: