Key result
Varying PVAD ejection timing alters wave interference, maximizing LV unloading during constructive interference.
Why the study?
Does the timing of pulsatile ventricular assist device ejection affect abdominal aortic wave intensity patterns and left ventricular unloading in a sheep model?
Does the timing of pulsatile ventricular assist device ejection affect abdominal aortic wave intensity patterns and left ventricular unloading in a sheep model?
Wave intensity analysis can assess the mechanical load of the left ventricle and peripheral hemodynamics during pulsatile VAD support, with optimal LV unloading occurring during constructive wave interference.
May support timing optimization to enhance LV unloading in pulsatile VADs; leaves open effects on clinical outcomes.
Arterial waves are seen as possible independent mediators of cardiovascular risks, and the wave intensity analysis (WIA) has therefore been proposed as a method for patient selection for ventricular assist device (VAD) implantation. Interpreting measured wave intensity (WI) is challenging, and complexity is increased by the implantation of a VAD. The waves generated by the VAD interact with the waves generated by the native heart, and this interaction varies with changing VAD settings. Eight sheep were implanted with a pulsatile VAD (PVAD) through ventriculoaortic cannulation. The start of PVAD ejection was synchronized to the native R wave and delayed between 0 and 90% of the cardiac cycle in 10% steps or phase shifts (PS). Pressure and velocity signals were registered, with the use of a combined Doppler and pressure wire positioned in the abdominal aorta, and used to calculate the WI. Depending on the PS, different wave interference phenomena occurred. Maximum unloading of the left ventricle (LV) coincided with constructive interference and maximum blood flow pulsatility, and maximum loading of the LV coincided with destructive interference and minimum blood flow pulsatility. We believe that noninvasive WIA could potentially be used clinically to assess the mechanical load of the LV and to monitor the peripheral hemodynamics such as blood flow pulsatility and risk of intestinal bleeding.
No takes yet. Share an insight, caveat, or question.
Jahren et al. (2014) studied this question. Pulsatile ventricular assist device (PVAD) timing was evaluated on Wave intensity pattern and interference phenomena. Varying the timing of pulsatile ventricular assist device ejection caused different wave interference phenomena, with maximum left ventricular unloading coinciding with constructive interference.