Key result
Active fixation atrial lead implantation results in delayed right atrial perforation successfully treated by extraction.
Case Report (n=1)
No
Highlights a rare but serious complication of right atrial perforation and hemopneumothorax following active fixation atrial lead implantation.
Answers are grounded in this paper and cite their sources.
Active fixation atrial leads warrant caution for delayed perforation; this case leaves open optimal lead selection in routine implantation.
A 24-year-old female underwent implantation of a right-sided dual chamber permanent pacemaker for cardioinhibitory syncope with active fixation atrial lead. Five days after the procedure, the patient developed pleuritic chest pain and difficulty in breathing. Minimal pericardial effusion and right hemopneumothorax were found. The atrial active screw-in lead was visualized just above the right atrial appendage with its helix perforating the right atrial wall, pericardium and pleura, reaching the right mid-lobe. Lead extraction was performed, and a passive bipolar atrial lead was implanted during the same session without any problems.
No takes yet. Share an insight, caveat, or question.
Avcı et al. (2013) conducted a case report in Cardioinhibitory syncope (n=1). Active fixation atrial lead implantation was evaluated on Right atrial perforation. Active fixation atrial lead implantation resulted in right atrial perforation, pericardial effusion, and hemopneumothorax 5 days post-procedure, successfully treated with lead extraction.
Synapse has enriched 10 related papers. Consider them for comparative context: