Key result
A left-sided dual chamber pacemaker implantation using a J-shaped active-fixation lead resulted in a contralateral right-sided pneumothorax due to right atrial wall perforation.
Case Report (n=1)
Highlights the risk of contralateral pneumothorax from atrial lead perforation and suggests passive or straight active-fixation leads may reduce perforation risk.
Answers are grounded in this paper and cite their sources.
Alerts to rare contralateral pneumothorax risk after atrial lead perforation; hypothesis-generating for lead design modifications.
We describe the occurrence of a right-sided pneumothorax following a left-sided dual chamber pacemaker implantation, due to a perforation of the screw of the J-shaped active-fixation lead through the right atrial wall. A review of the literature regarding complications of different atrial lead types (passive vs. active and J-shape vs. straight) is provided. Current data suggest that passive-fixation leads could have an advantage, provided they can be positioned satisfactorily because of the lower risk of pericardial complications. If active-fixation leads are used, straight leads might be preferable above J-shaped leads because of the lower rate of perforation.
No takes yet. Share an insight, caveat, or question.
Hugo Van Herendael (2009) conducted a case report in Contralateral pneumothorax after pacemaker implantation (n=1). J-shaped active-fixation atrial lead vs. Passive-fixation or straight active-fixation leads was evaluated on Perforation and pericardial complications. A left-sided dual chamber pacemaker implantation using a J-shaped active-fixation lead resulted in a contralateral right-sided pneumothorax due to right atrial wall perforation.
Synapse has enriched 10 related papers. Consider them for comparative context: