Key result
CoreValve linked to ~263% higher odds of post-dilatation vs SAPIEN XT despite similar paravalvular leak.
Why the study?
Does transcatheter aortic valve implantation with SAPIEN XT compared to CoreValve reduce paravalvular leak or need for post-dilatation in patients with severe aortic stenosis and severe device landing zone calcification?
Cohort (n=174)
No
Does transcatheter aortic valve implantation with SAPIEN XT compared to CoreValve reduce paravalvular leak or need for post-dilatation in patients with severe aortic stenosis and severe device landing zone calcification?
Answers are grounded in this paper and cite their sources.
In patients with severe device landing zone calcification undergoing TAVI, SAPIEN XT and CoreValve have similar rates of paravalvular leak, but SAPIEN XT requires less post-dilatation, though excessive oversizing with SAPIEN XT carries a risk of root injury.
Severe device landing zone calcification (DLZ-CA) predicted paravalvular leak (PVL) and post-dilatation (PD) after transcatheter aortic valve implantation (TAVI). The aim of this study was to determine the influence of DLZ-CA on PVL or PD rates after SAPIEN XT (XT) versus CoreValve (CV).We analyzed patients undergoing TAVI who had severe DLZ-CA. Severe DLZ-CA defined the upper left ventricular outflow tract calcification; the cross-sectional region 2 mm inferior to the annular plane. PVL was evaluated at 30days using transthoracic echocardiography. Overall, 133 patients had XT-TAVI and the remaining 41 patients had CV-TAVI. Two patients had annulus injury in the XT group (oversizing 20.2% and 20.5% for two XT cases). PD was less frequently performed in the XT group (34.1% versus 12.8%; P = 0.002), but PVL rates were similar between both groups (42.1% versus 41.5% for the XT and CV groups, respectively; P = 0.94). Importantly, excessive oversizing or the degree of filling volume was not associated with decreased PVL after XT-TAVI (P = non-significant for all). On multivariate analysis, CV-TAVI was found to be one of the independent predictors of need for PD (Odds ratio 3.63, 95% confidence interval 1.55 to 8.53, P = 0.003).In the setting of severe DLZ-CA, XT and CV have similar rates of PVL but XT had less need for PD. Excessive oversizing with XT carries a risk of root injury which could be further increased by DLZ-CA.
No takes yet. Share an insight, caveat, or question.
Maeno et al. (2017) conducted a cohort in Symptomatic severe aortic stenosis with severe device landing zone calcification (n=174). CoreValve vs. SAPIEN XT was evaluated on Need for post-dilatation (OR 3.63, 95% CI 1.55-8.53, p=0.003). CoreValve implantation in patients with severe device landing zone calcification predicted a higher need for post-dilatation compared to SAPIEN XT, with similar paravalvular leak rates.
Synapse has enriched 10 related papers. Consider them for comparative context: