Key result
Novel false lumen exclusion procedure achieves complete thrombosis and resolves hemothorax at 7 weeks.
Why the study?
Persistent retrograde false lumen flow is a common complication after TEVAR for type B aortic dissections, but device availability can limit the use of the candy plug technique.
Case Report (n=1)
An off-label technique using a Cook Converter graft and Amplatzer plugs offers a viable alternative for false lumen exclusion when standard candy plug devices are unavailable.
May offer an alternative for false lumen exclusion when dedicated devices are unavailable; hypothesis-generating and requires prospective validation.
Persistent retrograde false lumen (FL) flow is a common complication following thoracic endovascular aortic repair (TEVAR) for type B aortic dissections. The candy plug technique is a well-established method for FL exclusion; however, device availability can be a limiting factor. We report the case of a 57-year-old man who underwent a novel FL exclusion procedure 2 days post-TEVAR, using a 36 × 66 mm Cook Converter graft and Amplatzer plugs, with complete FL thrombosis and resolution of hemothorax on imaging 7 weeks postoperatively. This off-label technique offers a viable alternative when commercially available candy plug devices are unavailable and urgent intervention is required.
No takes yet. Share an insight, caveat, or question.
Crisp et al. (2025) conducted a case report in Retrograde false lumen flow causing hemothorax after TEVAR for type B aortic dissection (n=1). False lumen exclusion procedure using a Cook Converter graft and Amplatzer plugs was evaluated on Complete false lumen thrombosis and resolution of hemothorax on imaging. A novel false lumen exclusion procedure using a Cook Converter graft and Amplatzer plugs achieved complete false lumen thrombosis and resolution of hemothorax at 7 weeks postoperatively.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: