Key result
Digital RSA detects early glenoid component loosening more frequently than traditional radiographs.
Why the study?
Does digital roentgen stereophotogrammetric analysis (RSA) improve the detection of early loosening of cemented glenoid components compared to traditional radiographs in patients undergoing total shoulder arthroplasty?
Observational (n=48)
Does digital roentgen stereophotogrammetric analysis (RSA) improve the detection of early loosening of cemented glenoid components compared to traditional radiographs in patients undergoing total shoulder arthroplasty?
Traditional radiographs underestimate the rate of early loosening of cemented glenoid components in total shoulder arthroplasty compared to digital roentgen stereophotogrammetric analysis (RSA).
Answers are grounded in this paper and cite their sources.
RSA may improve early glenoid loosening detection; leaves open whether it should guide clinical decisions or revision timing.
The incidence of loosening of a cemented glenoid componentin total shoulder arthroplasty, detected by means of radiolucent lines or positional shift of the component on true anteroposterior radiographs, has been reported to be between 0% and 44%. Radiolucent lines are, however, difficult to detect and to interpret because of the mobility of the shoulder girdle and the obliquity of the glenoid which hinder standardisation of radiographs. We examined radiolucencies around cemented glenoid components in 48 patients, with a mean follow-up of 5.3 years, and found progressive changes to be present predominantly at the inferior pole of the component. This may hold a clue for the mechanism of loosening of this implant. In five patients we performed an additional analysis of loosening of the glenoid component using digital roentgen stereophotogrammetric analysis (RSA). After three years, three of the five implants had loosened (migration 1.2 to 5.5 mm). In only one, with gross loosening, were the radiological signs consistent with the RSA findings. When traditional radiographs are used for assessment, the rate of early loosening is underestimated. We recommend that RSA be used for this.
No takes yet. Share an insight, caveat, or question.
Nagels et al. (2002) conducted an observational in Total shoulder arthroplasty (n=48). Cemented glenoid component was evaluated on Radiolucencies and loosening of the glenoid component. Digital roentgen stereophotogrammetric analysis detected loosening in 3 of 5 cemented glenoid components after 3 years, indicating traditional radiographs underestimate early loosening.