Key result
A first-order time constant model of the heart demonstrates that for biphasic waveforms, a region of reliable low threshold defibrillation exists with phase 1/phase 2 durations from 3.5/1.5 to 9/6 ms.
A conceptual model of defibrillation identifies optimal phase durations for biphasic waveforms to achieve reliable low threshold defibrillation.
Answers are grounded in this paper and cite their sources.
Model suggests biphasic waveform optimization; leaves open prospective clinical validation before practice impact.
A model is developed for defibrillation that treats the heart as a first order time constant. Such a model allows ready evaluation of different monophasic waveforms. For implantable devices where the voltage is provided by the discharge of a capacitor, it can be seen that the effective voltage within the heart rises rapidly to a peak and then decays to zero. The time interval at which this peak occurs is defined as the optimal duration, and there is no advantage in extending the pulse beyond this point. Characteristics are presented that show how the time course of this voltage within the heart changes with different device capacitors and load impedances. The effect of different heart time constant are also examined. For biphasic waveforms, a contour plot of threshold voltage is presented with phase 1 and phase 2 durations on the two axes. It is seen that there is a region of reliable low threshold defibrillation from 3.5/1.5 ms to 9/6 ms.
No takes yet. Share an insight, caveat, or question.
B. G. Cleland (1996) studied Defibrillation. Defibrillation waveforms was evaluated on Threshold voltage for defibrillation. A first-order time constant model of the heart demonstrates that for biphasic waveforms, a region of reliable low threshold defibrillation exists with phase 1/phase 2 durations from 3.5/1.5 to 9/6 ms.
Synapse has enriched 10 related papers. Consider them for comparative context: