Key result
A new method of atrial evoked response detection was successful in all 16 patients during routine pacemaker implant, demonstrating that automatic atrial threshold tracking is feasible.
Why the study?
Does a new method of evoked response detection allow for reliable atrial capture detection and automatic threshold tracking in patients undergoing pacemaker implant?
Does a new method of evoked response detection allow for reliable atrial capture detection and automatic threshold tracking in patients undergoing pacemaker implant?
Answers are grounded in this paper and cite their sources.
A new method of evoked response detection allows for reliable atrial capture detection and makes automatic atrial threshold tracking feasible during pacemaker implantation.
A new method of evoked response detection, previously demonstrated in the ventricle, has been studied in the atrium at the time of routine pacemaker implant in 16 patients. The atrial evoked response was readily detectable in all patients due to excellent recovery from poststimulus polarization. In six patients, as experimental threshold-tracking pacemaker was used to automatically verify atrial capture and to generate strength-duration curves. It is concluded that this pacing technique is both simple and reliable, and that automatic atrial threshold tracking is feasible.
No takes yet. Share an insight, caveat, or question.
Livingston et al. (1988) studied Patients undergoing routine pacemaker implant (n=16). Atrial evoked response detection and experimental threshold-tracking pacemaker was evaluated on Detection of atrial evoked response. A new method of atrial evoked response detection was successful in all 16 patients during routine pacemaker implant, demonstrating that automatic atrial threshold tracking is feasible.