Key result
Myocardial viability loss in rabbit hearts reduces paced complex duration ~17% via shortened R-T periods.
The R-T shortening of the paced evoked potential is a sensitive measure of myocardial integrity in isolated rabbit hearts.
Answers are grounded in this paper and cite their sources.
May mark viability loss in rabbit models; leaves open clinical translation of paced-evoked responses.
The endocardial unipolar paced evoked response has excited a great deal of interest due to its possible use in the measurement of the metabolic state of the body and other pacer-related areas. Although rate-responsive pacing utilizing this signal has been clinically evaluated, little is known regarding the behavior of the components of this waveform under normal physiological conditions. We have developed an electronic circuit which allows the recording of the evoked response within a few milliseconds of a pacing stimulus of 5 V and 0.5 ms duration being applied using a single unipolar, smooth platinum electrode of 14 mm2 surface area. The paced evoked response was measured using a total of 20 isolated rabbit heart preparations. Five were run for 8 hours and the remaining fifteen were run for 5 hours. Our results indicate that the waveform components of the evoked response remain stable while the preparation is viable, but that two of the time-related measurements change with loss of viability. A significant lengthening of the stimulus-R interval was seen together with a dramatic shortening of the R-T period. The net result of these changes was an overall reduction of 17% in the complex duration. In addition, we found the R-T shortening to be a sensitive measure of myocardial integrity. We conclude that the combination of our interface charge elimination circuit and the isolated heart preparation has proved a useful system for the investigation of the paced evoked potential. Furthermore, the loss of myocardial viability has a complex action on this response.
No takes yet. Share an insight, caveat, or question.
Economides et al. (1988) studied Isolated animal heart (n=20). Pacing stimulus using a single unipolar platinum electrode was evaluated on Waveform components of the paced evoked response (stimulus-R interval, R-T period, complex duration) (17% reduction). Loss of myocardial viability in isolated rabbit hearts resulted in a significant lengthening of the stimulus-R interval and a dramatic shortening of the R-T period, reducing complex duration by 17%.
Synapse has enriched 10 related papers. Consider them for comparative context: