Key result
Reflotron dry-chemistry measures total cholesterol with a ~0.22 mmol/L difference versus standard Hitachi analyzers.
Why the study?
Does the Reflotron dry-chemistry system accurately and reproducibly measure plasma total cholesterol levels compared to standard laboratory methods in volunteers?
Cross-Sectional (n=80)
Does the Reflotron dry-chemistry system accurately and reproducibly measure plasma total cholesterol levels compared to standard laboratory methods in volunteers?
The Reflotron dry-chemistry system provides rapid, accurate, and reproducible total cholesterol measurements suitable for community screening.
Answers are grounded in this paper and cite their sources.
Supports feasibility for community screening; leaves open prospective validation before routine use.
This study compared the measurements of total cholesterol levels by means of a Reflotron dry-chemistry system and a standard laboratory method with the Hitachi 705 analyser. Simultaneous blood samples were collected by finger-prick and by venepuncture from 80 volunteers. The ranges of total cholesterol levels were 3.38-8.45 mmol/L (Reflotron) and 3.59-8.83 mmol/L (Hitachi). The mean (+/- standard deviation [SD]) difference between the two methods was 0.22 +/- 0.30 mmol/L. All the subjects provided a second finger-prick sample of blood on the next day to assess the reproducibility of the Reflotron system. The mean (+/- SD) cholesterol levels that were obtained by the Reflotron method on Day 1 and Day 2 were 5.38 +/- 1.17 mmol/L and 5.30 +/- 1.14 mmol/L, respectively. These results did not differ significantly (mean difference, 0.079 mmol/L; standard error, 0.045 mmol/L). Provided that the operator is trained appropriately, the Reflotron system gives rapid, accurate and reproducible results.
No takes yet. Share an insight, caveat, or question.
Phillips et al. (1988) conducted a cross-sectional in Cholesterol screening (n=80). Reflotron dry-chemistry system vs. Hitachi 705 analyser was evaluated on Difference in total cholesterol levels between methods (MD 0.22). The Reflotron dry-chemistry system measured total cholesterol levels with a mean difference of 0.22 mmol/L compared to the standard Hitachi 705 analyser.
Synapse has enriched 10 related papers. Consider them for comparative context: