Key result
Capillary point-of-care total cholesterol yields similar results to venous sampling.
Why the study?
The relationship between cholesterol determinations in venous and capillary samples using the CR3000 PoC system was not well established.
Does capillary blood sampling provide similar cholesterol measurements compared to venous sampling using the CR3000 PoC system in adult volunteers?
Observational (n=21)
Does capillary blood sampling provide similar cholesterol measurements compared to venous sampling using the CR3000 PoC system in adult volunteers?
The CR3000 point-of-care system provides comparable and accurate total cholesterol measurements using either capillary or venous blood samples.
Answers are grounded in this paper and cite their sources.
Comparable results support capillary sampling flexibility with CR3000; leaves open prospective validation before routine adoption.
Currently, the clinical practice of desktop or Point of Care (PoC) analyzers for lipid measurements has gained wide popularity. Designed to quickly perform measurements on microlitre(microL) quantities of blood, these instruments can be used in non-laboratory settings, such as physicians offices or field-testing sites and can provide measurements in whole blood, serum, or plasma, using either venous or capillary blood samples. The aim of this study is to examine the relationship between cholesterol determinations in venous and capillary samples using the CR3000 PoC system. The study was performed on 21 unselected adult volunteers, and no exclusion criteria was adopted. The mean cholesterol concentration for the venous blood samples measured was 164 mg/dL. The values obtained in the capillary blood samples averaged 168 mg/dL, which is only slightly higher (e.g., 2.87%) than the venous sample measurements. Moreover, the total variance was statistically similar for venous and capillary measurements (F value = 1.199, where the upper critical value of the F distribution is 2.124, p < 0.05). The results of our study support the concept that CR3000 total cholesterol testing can be performed safely and accurately in either venous or capillary specimens.
No takes yet. Share an insight, caveat, or question.
Sblendorio et al. (2008) reported an observational. Capillary blood sampling using CR3000 vs. Venous blood sampling using CR3000 was evaluated on Total cholesterol concentration (p=<0.05). Capillary blood sampling for total cholesterol using the CR3000 system yielded similar results to venous sampling, with similar variance.