Alternating-current swimming pool leaks likely trigger 2 inappropriate shocks in a biventricular ICD patient.
Case Report (n=1)
Alternating-current leaks in swimming pools can mimic ventricular fibrillation and cause inappropriate ICD shocks.
Answers are grounded in this paper and cite their sources.
Pool electrical hazards warrant counseling for ICD patients; this case leaves open the incidence of AC-triggered shocks.
Implantable cardioverter-defibrillators (ICDs) are the standard of care for preventing sudden cardiac death in patients who are predisposed to malignant ventricular arrhythmias. Causes of inappropriate ICD shock include equipment malfunction, improper arrhythmia evaluation, misinterpretation of myopotentials, and electromagnetic interference. As the number of implanted ICDs has increased, other contributors to inappropriate therapy have become known, such as minimal electrical current leaks that mimic ventricular fibrillation. We present the case of a 63-year-old man with a biventricular ICD who received 2 inappropriate shocks, probably attributable to alternating-current leaks in a swimming pool. In addition, we discuss ICD sensitivity and offer recommendations to avoid similar occurrences.
No takes yet. Share an insight, caveat, or question.
Makaryus et al. (2014) conducted a case report in Inappropriate ICD shock (n=1). Alternating-current leak in a swimming pool was evaluated on Inappropriate ICD shock. Alternating-current leaks in a swimming pool likely caused 2 inappropriate shocks in a 63-year-old man with a biventricular ICD.