Key result
Standard stun gun shocks do not affect pacemaker or ICD function in a swine model.
Why the study?
Does a standard 5-second electrical stun gun shock affect the functional integrity of implantable pacemakers and defibrillators in a swine model?
Does a standard 5-second electrical stun gun shock affect the functional integrity of implantable pacemakers and defibrillators in a swine model?
Standard 5-second electrical stun gun shocks do not compromise the functional integrity of implantable pacemakers and defibrillators in a swine model.
AIMS: High voltage electric current can adversely affect pacemakers (PM) and implantable cardioverter-defibrillator (ICD). The standard shock from an electrical stun gun (TASER- X26, TASER International, Scottsdale, AZ) consists of a 5-s long application of high voltage, low current pulses at 19 pulses per second. Its effect on the functional integrity of PM and ICDs is unknown. METHODS AND RESULTS: We tested the functional integrity of nine PMs and seven ICDs in a swine model after a standard stun gun shock. A transvenous, dual coil, bi-polar ICD lead (St Jude-SP01) and a PM lead were placed in the right ventricular (RV) apex and connected to pulse generators buried in the pre-pectoral pocket. The two darts were placed at the sternal notch (SN) and apex of the heart bracketing the device pocket. Standard neuromuscular incapacitating (NMI) discharges were delivered. Functional parameters of the devices and leads were checked before and after the shocks. The mean pacing thresholds, sensing thresholds, pacing impedances, and defibrillation coil impedances of the ICD lead were similar before and after the shocks. Similarly, pacing thresholds, sensing thresholds, and impedances of the PM lead were not significantly different before and after the shocks. No significant change was noted in battery voltage and projected longevity. Implantable cardioverter-defibrillator generators detected the NMI impulses at a mean cycle length of 176 +/- 20 ms with detection to charge time of 5.9 +/- 1.5 s. Shock delivery was aborted in all tests as tachycardia detection abruptly terminated at the end of the 5 s NMI application. None of the devices exhibited power on reset (POR), elective replacement indicator (ERI), or noise mode behaviour after the shock. CONCLUSION: Pacemakers and ICD generators and leads functions were not affected by the tested standard 5 s stun gun shocks.
No takes yet. Share an insight, caveat, or question.
Khasnis et al. (2007) studied Implantable pacemakers and defibrillators (n=16). Electrical stun gun (TASER-X26) shock vs. Baseline (before shock) was evaluated on Functional integrity of pacemakers and ICDs (pacing thresholds, sensing thresholds, impedances). Standard 5-second stun gun shocks did not affect pacemaker or ICD function in a swine model, though ICDs detected impulses at a mean cycle length of 176 +/- 20 ms.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: