Key result
Transcutaneous electrical nerve stimulation caused interference with implantable cardioverter defibrillators in 16 of 30 patients, including misinterpretation as VT/VF in 27%.
Why the study?
Does transcutaneous electrical nerve stimulation (TENS) cause interference with the sensing function of implantable cardioverter defibrillators (ICDs)?
Observational (n=30)
Does transcutaneous electrical nerve stimulation (TENS) cause interference with the sensing function of implantable cardioverter defibrillators (ICDs)?
TENS causes significant electromagnetic interference with ICD sensing, potentially leading to inappropriate shocks or failure to deliver therapy, and is therefore not recommended in patients with ICDs.
Answers are grounded in this paper and cite their sources.
TENS may disrupt ICD sensing; this observational finding is hypothesis-generating and requires prospective validation before practice change.
BACKGROUND: The use of transcutaneous electrical nerve stimulation (TENS) for pain relief is increasing. At the same time the implantable cardioverter defibrillator (ICD) is a routine treatment for malignant tachyarrhythmias. Today patients often need devices for more than one condition, and consideration must be given to the interaction between them. We studied the risk of interference between TENS and the ICD function. METHODS AND RESULTS: Thirty patients who had received an ICD underwent a test protocol including TENS at the mammilla and hip levels, at two energy levels, and at the highest comfortable stimulation level. The effects of TENS on the electrocardiogram lead II, intracardiac electrograms, and the ICD marker channels were analyzed. Disturbance from TENS on the sensing function was seen at all stimulation attempts. Interference between the systems was observed in 16 patients. In eight patients (27%) the interpretation was VT/VF and in 14 patients (47%) as ventricular premature extra beats. Other kinds of interactions were seen in five patients (16%). Each patient could have more than one kind of interference. CONCLUSIONS: Noise reversion and undersensing might prevent the ICD from delivering shock when it should and the interpretation as VT/VF could result in inappropriate shocks. Because of the potentially serious consequences of interference we do not recommend the use of TENS in patients with ICD.
No takes yet. Share an insight, caveat, or question.
Holmgren et al. (2008) conducted an observational in Implantable cardioverter defibrillator (ICD) (n=30). Transcutaneous electrical nerve stimulation (TENS) was evaluated on Interference between TENS and ICD function. Transcutaneous electrical nerve stimulation caused interference with implantable cardioverter defibrillators in 16 of 30 patients, including misinterpretation as VT/VF in 27%.
Synapse has enriched 10 related papers. Consider them for comparative context: