TENS unit use triggers inappropriate ICD shocks via electrical artifact mimicking VF.
Case Report (n=1)
TENS units can interfere with transvenous ICDs by creating electrical artifacts that trigger inappropriate shocks, warranting caution in this patient population.
Answers are grounded in this paper and cite their sources.
TENS units may trigger inappropriate ICD shocks; this case report leaves open the frequency and predictors of electromagnetic interference.
Although the potential for transcutaneous electronic nerve stimulation (TENS) units to interfere with appropriate function of cardiac pacemakers is well documented, an adverse interaction between a TENS unit and an ICD has never been reported. We describe a patient in whom a TENS unit created an electrical artifact that was interpreted by a transvenous ICD as ventricular fibrillation, leading to the delivery of inappropriate therapy. TENS units should be used with caution in patients with ICDs.
No takes yet. Share an insight, caveat, or question.
PHILBIN et al. (1998) conducted a case report in Patient with an implantable cardioverter-defibrillator (ICD) (n=1). Transcutaneous electronic nerve stimulation (TENS) unit was evaluated on Inappropriate therapy (shocks) due to electrical artifact interpreted as ventricular fibrillation. Use of a TENS unit in a patient with a transvenous ICD created an electrical artifact interpreted as ventricular fibrillation, leading to the delivery of inappropriate shocks.