Key result
Transcutaneous electrical nerve stimulation (TENS) applied to the thoracic and lumbar regions produced an electrocardiographic artifact resembling a malfunctioning cardiac pacemaker spike.
Case Report (n=1)
Physicians should be aware that TENS units can produce EKG artifacts resembling pacemaker spikes to avoid misdiagnosing pacemaker malfunction.
Answers are grounded in this paper and cite their sources.
TENS may generate ECG artifacts mimicking pacemaker malfunction; case report leaves open broader awareness needs.
Transcutaneous electrical nerve stimulation (TENS) is an effective modality for the treatment of pain. TENS use is well tolerated and has relatively few complications. Potentially serious complications of TENS use include disruption of cardiac pacemaker function and electrical artifact during electrocardiographic monitoring. Electrical interference with cardiac monitoring using parasternal electrodes and electrocardiographic (EKG) artifact with an epidural stimulator has been reported. We report the case of EKG artifact produced by surface TENS electrodes applied to the thoracic and lumbar regions. The electrical signal produced by the TENS unit created an electrocardiographic artifact resembling a pacemaker spike on routine EKG. This was interpreted as a malfunctioning cardiac pacemaker. The easily misinterpreted electrical spikes that can result from TENS use as seen in this case highlights the need for physician awareness of this potential complication.
No takes yet. Share an insight, caveat, or question.
James A. Sliwa (1996) conducted a case report in Electrocardiogram artifact (n=1). Transcutaneous electrical nerve stimulation (TENS) was evaluated on Electrocardiographic artifact resembling a pacemaker spike. Transcutaneous electrical nerve stimulation (TENS) applied to the thoracic and lumbar regions produced an electrocardiographic artifact resembling a malfunctioning cardiac pacemaker spike.
Synapse has enriched 10 related papers. Consider them for comparative context: