TENS units cause ECG artifacts on Holter monitoring that simulate atrial flutter.
Why the study?
Artifact caused by transcutaneous electrical nerve stimulation (TENS) units can simulate atrial flutter on Holter monitoring, potentially leading to misdiagnosis.
TENS units can cause ECG artifacts that mimic atrial flutter on Holter monitoring, highlighting the need for careful interpretation to avoid misdiagnosis.
Answers are grounded in this paper and cite their sources.
TENS-related Holter artifacts warrant consideration before diagnosing atrial flutter; leaves open systematic assessment of prevalence and recognition strategies.
A patient with previous actual atrial flutter had what appeared to be atrial flutter seen on Holter monitoring during treatment with a transcutaneous electrical nerve stimulation (TENS) unit. This proved to be artifact caused by the unit rather than a true arrhythmia. The artifact was reproduced in a normal volunteer by application of a TENS unit.
No takes yet. Share an insight, caveat, or question.
Weitz et al. (1997) studied this question. Transcutaneous electrical nerve stimulation (TENS) units can cause electrocardiographic artifacts on Holter monitoring that simulate atrial flutter.