Key result
Multi-wavelength optoelectronic patch sensor shows strong agreement with SpO2 during rest and exercise.
Why the study?
Does a multi-wavelength optoelectronic patch sensor using green and orange illuminations provide accurate SpO2 measurements during movement in healthy subjects?
Observational (n=31)
Does a multi-wavelength optoelectronic patch sensor using green and orange illuminations provide accurate SpO2 measurements during movement in healthy subjects?
Answers are grounded in this paper and cite their sources.
A novel optoelectronic patch sensor using green and orange illuminations provides robust SpO2 measurements even during physical movement.
Photoplethysmography (PPG) based pulse oximetry devices normally use red and infrared illuminations to obtain oxygen saturation (SpO2) readings. In addition, the presence of motion artefacts severely restricts the utility of pulse oximetry physiological measurements. In the current study, a combination of green and orange illuminations from a multi-wavelength optoelectronic patch sensor (mOEPS) was investigated in order to improve robustness to subjects’ movements in the extraction of SpO2 measurement. Two experimental protocols with 31 healthy subjects were designed to determine SpO2 measurement. The datasets for the first protocol were collected from 15 subjects at rest, with the subjects free to move their hands. The datasets for the second protocol with 16 subjects were collected during cycling and running exercises. The results showed good agreements with SpO2 measurements (r = 0.98) in the both protocols. The outcomes promise a robust and cost-effective approach of physiological monitoring with the prospect of providing health monitoring that does not restrict user physical movements.
No takes yet. Share an insight, caveat, or question.
Alharbi et al. (2018) conducted an observational in Healthy (n=31). Multi-wavelength optoelectronic patch sensor (mOEPS) with green and orange illuminations was evaluated on SpO2 measurement agreement (r = 0.98). A multi-wavelength optoelectronic patch sensor using green and orange illuminations showed good agreement with SpO2 measurements (r = 0.98) in healthy subjects at rest and during exercise.