This is the authors' abstract. We don't add key points for this paper.
A fully implantable neural data acquisition system is a key component of a clinically viable cortical brain-machine interface. We present the design and implementation of a single-chip device that serves the processing needs of such a system. Our device processes 96 channels of multi-unit neural data and performs all digital processing necessary for bidirectional wireless communication. The implementation utilizes a single programmable logic device that is responsible for performing data reduction on the 96 channels of neural data, providing a bidirectional telemetry interface to a transceiver and performing command interpretation and system supervision. The device takes as input neural data sampled at 31.25 kHz and outputs a line-encoded serial bitstream containing the information to be transmitted by the transceiver. Data can be output in one of the following four modes: (1) streaming uncompressed data from a single channel, (2) extracted spike waveforms from any subset of the 96 channels, (3) 1 ms bincounts for each channel or (4) streaming data along with extracted spikes from a single channel. The device can output up to 2000 extracted spikes per second with latencies suitable for a brain-machine interface application. This device provides all of the digital processing components required by a fully implantable system.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Rizk et al. (2007) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: