Key result
Novel two-electrode ECG amplifier yields ~0.5% residual power vs standard three-electrode circuits.
Answers are grounded in this paper and cite their sources.
A novel two-electrode biosignal amplifier achieves high signal quality comparable to standard three-electrode ECGs while consuming only 11 μW of power.
A novel two-electrode biosignal amplifier circuit is demonstrated by using a composite transimpedance amplifier input stage with active current feedback. Micropower, low gain-bandwidth product operational amplifiers can be used, leading to the lowest reported overall power consumption in the literature for a design implemented with off-the-shelf commercial integrated circuits (11 μW). Active current feedback forces the common-mode input voltage to stay within the supply rails, reducing baseline drift and amplifier saturation problems that can be present in two-electrode systems. The bandwidth of the amplifier extends from 0.05-200 Hz and the midband voltage gain (assuming an electrode-to-skin resistance of 100 kΩ) is 48 dB. The measured output noise level is 1.2 mV pp, corresponding to a voltage signal-to-noise ratio approaching 50 dB for a typical electrocardiogram (ECG) level input of 1 mVpp. Recordings were taken from a subject by using the proposed two-electrode circuit and, simultaneously, a three-electrode standard ECG circuit. The residual of the normalized ensemble averages for both measurements was computed, and the power of this residual was 0.54% of the power of the standard ECG measurement output. While this paper primarily focuses on ECG applications, the circuit can also be used for amplifying other biosignals, such as the electroencephalogram.
No takes yet. Share an insight, caveat, or question.
Inan et al. (2009) studied this question. Two-electrode transimpedance biosignal amplifier with active current feedback stabilization vs. Three-electrode standard ECG circuit was evaluated on Power of the residual of the normalized ensemble averages between the two measurements. A novel two-electrode biosignal amplifier circuit demonstrated a residual power of 0.54% compared to a standard three-electrode ECG circuit, with an overall power consumption of 11 μW.
Synapse has enriched 10 related papers. Consider them for comparative context: