Key result
Simulated ten-minute EHR time deviations increase pulse oximetry error versus lab measurements by ~114%.
Why the study?
Studies using electronic health record data to evaluate race-based pulse oximetry imprecision do not appear to account for temporal variability in SpO2 values.
Answers are grounded in this paper and cite their sources.
EHR data cannot be reliably used to quantify SpO2 error due to the impact of time deviation variability, suggesting caution in interpreting prior studies that used such data.
Purpose Articles utilizing large data sets from electronic health records (EHR) have been used to demonstrate race-based imprecision in pulse oximetry (SpO2) measurements. These articles do not appear to recognize the impact of the variability of the SpO2 values with respect to time (“time deviation”). This manuscript seeks to demonstrate that due to this variability, EHR data should not be used to quantify SpO2 error Methods Utilizing the MIMIC-IV Waveform dataset, SpO2 values are sampled and used as reference samples. The error derived from the EHR data is simulated using a set of time deviations. The laboratory oxygen saturation measurements are also simulated using an average root mean squared (ARMS) error of 2%. The analysis is then undertaken to quantify the mean error, the standard deviation of the error, the ARMS error, and the Bland-Altman Limits of Agreements. Results All measured parameters appeared to increase as a function of the logarithm of the deviation time. At 10 minutes, the ARMS error increased from a baseline of 2% to over 4% Conclusion EHR data cannot be reliably used to quantify SpO2 error. Caution should be used in interpreting prior manuscripts.
No takes yet. Share an insight, caveat, or question.
Élie Sarraf (2023) studied Pulse oximetry measurement error (n=198). Time deviation between EHR-logged SpO2 and laboratory measurement vs. Baseline (no time deviation) was evaluated on Average Root Mean Squared (ARMS) error at 10 minutes (95% CI 4.06-4.49, p=<1e-5). A simulated 10-minute time deviation between EHR-logged pulse oximetry and laboratory measurements increased the Average Root Mean Squared error from a baseline of 2% to over 4%.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: