Key result
Matching respiratory and cardiac cycles during CartoSound-CT integration is linked to ~32% less integration error.
Why the study?
Does matching respiratory phase and cardiac cycle improve the accuracy of 3D left atrial CartoSound and CT image integration in atrial fibrillation patients undergoing pulmonary vein isolation?
Cohort (n=56)
Does matching respiratory phase and cardiac cycle improve the accuracy of 3D left atrial CartoSound and CT image integration in atrial fibrillation patients undergoing pulmonary vein isolation?
Matching respiratory phase and cardiac cycle between CartoSound and CT image acquisition significantly improves the accuracy of 3D left atrial image integration during pulmonary vein isolation.
Answers are grounded in this paper and cite their sources.
May improve image integration accuracy in AF ablation; hypothesis-generating and requires prospective validation before practice change.
BACKGROUND: CartoSound™ (CS) module is useful in integrating 3-dimensional (3D) left atrial (LA) image with CT image. Integration method, however, has not been established. We reported the accuracy of LA electroanatomical (EA) and CT image integration by registering LA roof (LAR) and posterior wall (LAPW). METHODS: The consecutive 56 atrial fibrillation patients undergoing pulmonary vein isolation were studied. In the initial 29 patients, before the transseptal puncture, 3D CS LAR and LAPW image was created by registering a mean of 10 contour lines between the right and left pulmonary veins. After transseptally inserting a mapping catheter into LA, 3D EA image of LAR and LAPW was obtained by sampling a mean of 40 points. LA CT image was taken at the full-inspiratory position and 0% of R-R interval. After visual alignment of CS or EA and LA CT image, the 2 images were integrated with surface registration program. In the latter 27 patients, both CT and CS images were taken while matching the respiratory phase at the end-tidal position and cardiac cycle at 50% of R-R interval. RESULTS: In the initial 29 patients, the mean distances between EA and CT images and between CS and CT images were 1.53 ± 0.27 and 1.59 ± 0.23 mm, respectively (P = NS). In the latter 27, the mean distance was decreased to 1.08 ± 0.14 mm (P < 0.0001). CONCLUSIONS: CS system is useful in image integration with 3D CT. Matching both respiratory phase and cardiac cycle between CS and CT image acquisition improves the image integration accuracy.
No takes yet. Share an insight, caveat, or question.
Kimura et al. (2013) conducted a cohort in Atrial fibrillation (n=56). Matching respiratory phase and cardiac cycle during image acquisition vs. Image acquisition without matching respiratory phase and cardiac cycle was evaluated on Mean distance between CartoSound and CT images (p=<0.0001). Matching respiratory phase and cardiac cycle between CartoSound and CT image acquisition improved integration accuracy, reducing the mean distance from 1.59 mm to 1.08 mm.