Answers are grounded in this paper and cite their sources.
Due to the fact that the risk of Ni‐allergies are more and more important for modern therapies, the need of Ni‐free implant materials becomes increasingly vital. Beside Co‐ and Ti‐base alloys the new Ni‐free high‐Nitrogen steels might offer an attractive alternative to these expensive materials. The high‐Nitrogen and Ni‐free stainless steel X13CrMnMoN18‐4‐3 (1.4452) is investigated in the solution annealed state as to its corrosion, fatigue and sliding wear behaviour in Ringer solution. Under all test stresses this steel revealed a better behaviour as the conventional 316L‐type X2CrNiMo18‐15‐3. When compared to literature data of solution annealed Ni‐containing high‐Nitrogen steels and Ni‐free Co‐base alloys the 1.4452 shows similar bahaviour. The good fatigue behaviour is attributed to the low stacking fault energy resulting in the total absence of climbing and cross slip. The corrosion properties can be related to the amount of dissolved N. The sliding wear behaviour in a self‐mating contact benefits from both the low stacking fault energy and the characteristic corrosion mechanisms. Together with the elsewhere reported non‐cytotoxicity the steel X13CrMnMoN18‐4‐3 (1.4452) can be considered being suitable for medical applications.
No takes yet. Share an insight, caveat, or question.
Koch et al. (2002) studied this question.