Key result
STIC power Doppler reduces vascularization flow index variance vs conventional static 3D power Doppler.
Why the study?
Does STIC power Doppler reduce the variability of vascularization flow index compared to static 3D power Doppler in flow phantom and carotid artery models?
Does STIC power Doppler reduce the variability of vascularization flow index compared to static 3D power Doppler in flow phantom and carotid artery models?
Answers are grounded in this paper and cite their sources.
3D power Doppler indices derived from STIC are more stable and repeatable than those obtained from conventional static 3D power Doppler datasets.
OBJECTIVES: To compare the variability in vascularization flow index (VFI) seen in serial acquisitions obtained using spatiotemporal image correlation (STIC) and using conventional static three-dimensional (3D) power Doppler (PD), for both in-vitro and in-vivo models, and to evaluate whether the curves formed by VFI values obtained from successive 'frames' in a STIC dataset are consistent and resemble the waveforms obtained by spectral Doppler analysis. METHODS: The study was divided into two parts: in the first part (the in-vitro model) we scanned a flow phantom, while in the second part (the in-vivo model) we scanned a common carotid artery. Conventional static 3D and STIC-PD datasets were alternately acquired from these two models. VFI values were assessed from 0.38-cm(3) spherical samples of the main flow region in the static 3D datasets and in every frame of the STIC datasets. The variance of the minimum, mean and maximum VFI values from each STIC dataset was compared with the variance of VFI values from the static 3D datasets. RESULTS: Ten static 3D and 10 STIC datasets were acquired from each model. Analysis of the in-vitro and in-vivo models showed a significant reduction in the variance of VFI values obtained using STIC as compared to static datasets. Additionally, we observed that the curves formed by VFI values obtained from successive frames in each STIC dataset were consistent across different datasets and that they resembled the waveforms obtained by spectral Doppler in both models. CONCLUSIONS: 3D-PD indices derived from STIC are more stable than those obtained from conventional static 3D-PD datasets. The curves of VFI throughout a reconstructed cardiac cycle using STIC are repeatable and resemble those obtained by spectral Doppler analysis of the vessel.
No takes yet. Share an insight, caveat, or question.
Miyague et al. (2013) studied this question. Spatiotemporal image correlation (STIC) power Doppler vs. Conventional static 3D power Doppler was evaluated on Variance of vascularization flow index (VFI) values. Spatiotemporal image correlation (STIC) power Doppler significantly reduced the variance of vascularization flow index values compared to conventional static 3D power Doppler.