Key result
Using the RV anode cuts monophasic defibrillation energy by ~33% but shows no biphasic benefit.
Why the study?
Does electrode polarity affect the energy required for successful defibrillation using monophasic and biphasic waveforms in a porcine model?
Does electrode polarity affect the energy required for successful defibrillation using monophasic and biphasic waveforms in a porcine model?
In a porcine model, using the RV electrode as an anode reduces the energy required for defibrillation with monophasic shocks, but polarity does not affect biphasic shock efficacy.
Answers are grounded in this paper and cite their sources.
Polarity may lower monophasic defibrillation energy in pigs; hypothesis-generating and should not change clinical practice.
Electrode polarity has been reported to be one of the factors that affect defibrillation efficacy. We studied the influence of polarity on defibrillation efficacy when monophasic and biphasic waveforms were used with an endocardial lead system. In six anesthetized pigs, defibrillation catheters were placed in the right ventricular (RV) apex and at the junction of the superior vena cava (SVC) and right atrium. Monophasic shocks were 6 ms in duration, while for biphasic shocks the first phase was 6 ms and the second was 4 ms in duration. Four electrode configurations were tested: R:S, M (the RV electrode, cathode; the SVC electrode, anode, with a monophasic shock); S:R: S, B(the RV electrode, first phase cathode; the SVC electrode, first phase anode, with a biphasic shock); S:R, B. Defibrillation probability of success curves were determined using an up/down protocol requiring 15 shocks for each configuration. For monophasic shocks, total delivered energy at the 50% probability of success point was significantly lower when the RV electrode was an anode than when it was a cathode (R: S, M: 24.4 +/- 7.4 J [mean +/- SD] vs S:R, M: 16.4 +/- 5.5 J; P < 0.05). For biphasic shocks, total energy was not affected by polarity reversal of the electrodes (R:S, B: 8.7 +/- 1.4 J vs S:R, B: 8.4 +/- 2.5 J; P = NS). The endocardial electrode configuration with the RV electrode as an anode requires less energy for defibrillation with a monophasic but not a biphasic waveform.
No takes yet. Share an insight, caveat, or question.
Usui et al. (1996) studied this question. Right ventricular electrode as anode vs. Right ventricular electrode as cathode was evaluated on Total delivered energy at the 50% probability of success point for monophasic shocks (p=<0.05). Using the right ventricular electrode as an anode required less energy for defibrillation with a monophasic waveform, but not with a biphasic waveform.