Key result
Ultra-wideband radar accurately tracks heart rate in moving persons with ~0.97 correlation versus PPG.
Why the study?
Does a 79 GHz ultra-wideband radar sensor accurately monitor instantaneous heart rate in moving persons compared to finger PPG?
Does a 79 GHz ultra-wideband radar sensor accurately monitor instantaneous heart rate in moving persons compared to finger PPG?
A 79 GHz ultra-wideband radar sensor with advanced signal processing can potentially monitor instantaneous heart rate in moving individuals.
May support contactless HR monitoring during motion; leaves open prospective validation before clinical adoption.
This paper presents a 79 GHz ultra-wideband sensor monitoring the instantaneous heart-rate (HR) of moving persons. The sensor uses three approaches for the instantaneous HR monitoring: block-based motion/movement compensation of multiple persons, multiresolution analysis and Burg method for instantaneous HR analysis. The first approach normalizes and/or compensates for motion/movement of moving persons where the consecutive received signal, called radar range-profile, is stored on a block-by-block basis and the motion/movement within each block is then normalized as stationary. The second approach uses multi-resolution analysis (MRA) in order to remove most of the breathing signal from the received vital-sign waveform (HR and breathing). And the third approach estimates the instantaneous HR using Burg method. Measurement was conducted in order to investigate the usefulness of our suggested HR monitoring system including the above three approaches. The estimated HR is also compared with an optical pulse wave sensor (finger PPG) in order to confirm the accuracy.
No takes yet. Share an insight, caveat, or question.
Shibao et al. (2020) studied Healthy (heart rate monitoring) (n=3). 79 GHz ultra-wideband radar sensor vs. Optical pulse wave sensor (finger PPG) was evaluated on Correlation coefficient of heart rate intervals compared to PPG (Correlation coefficient 0.96-0.98). A 79 GHz ultra-wideband radar sensor accurately estimated instantaneous heart rate in moving persons, demonstrating a correlation coefficient of 0.96 to 0.98 when compared to a finger PPG sensor.