Answers are grounded in this paper and cite their sources.
Microcantilever-based biosensors have shown the ability of highly sensitive label-free detection by allowing the transduction of a target mass into a resonant frequency shift. Even if they have been widely adopted in the last years as high sensitive mass detectors, precision and specificity of such tools are often inadequate. A new platform able to precisely characterize in dynamic mode several of these micromechanical resonators is here presented. Consequently, to the examination of the main parameters involved in a measurement, repeatability and reproducibility of the automated cantilever-based system were estimated. The reported analyses demonstrated that the uncertainty of the system could be reduced to a standard deviation of 6.5 × 10-6(parts per million). Thanks to its fine precision, the here-proposed sensing platform was able to detect aflatoxins in naturally contaminated nuts at concentrations as low as 40 pg/mL, nearly two orders of magnitude more sensitive than previous attempts reported in the literature with similar devices.
No takes yet. Share an insight, caveat, or question.
Ferrante et al. (2016) studied this question.