Answers are grounded in this paper and cite their sources.
OBJECTIVE: The purpose of this study was to determine the accuracy of noninvasive hemoglobin measurement using pulse CO-oximetry (Pronto® SpHb®, Masimo Corp.) and a commonly used, invasive, point-of-care, automated spectrophotometer (HemoCue 201+®, HemoCue, Inc.), in comparison with hemoglobin measurements obtained from a laboratory hematology analyzer (measuring total blood hemoglobin) in the outpatient setting. METHODS: Adult patients presenting to an outpatient research clinic were tested for total blood hemoglobin measurement by 3 methods: noninvasive pulse CO-oximetry (SpHb®), finger-stick blood sample on a point-of-care device, and venous sample on a laboratory hematology analyzer (reference device). Bias and standard deviation (SD) of SpHb® and HemoCue 201+® compared with the values obtained with the laboratory hematology analyzer were calculated and Bland-Altman graphs were generated. RESULTS: Samples from 152 subjects were assessed (average age, 46 years; 69% female). The bias ± SD compared with the reference method was -0.5 ± 1.0 g/dL for SpHb® and 0.3 ± 1.0 g/dL for HemoCue 201+®. The Bland-Altman plots assessing agreement of the test methods to the reference method had limits of agreement of -2.5 to 1.5 g/dL for SpHb® and -1.7 to 2.3 g/dL for HemoCue 201+®. A noninvasive measurement could not be obtained in 4 subjects after 2 attempts (2.5% failure rate), whereas the HemoCue 201+® measurements were obtained for all subjects. CONCLUSION: Noninvasive SpHb® testing had bias and SD similar to those of HemoCue 201+®. Because SpHb® measurement is noninvasive, it may offer additional patient and provider benefits.
No takes yet. Share an insight, caveat, or question.
Marina Raikhel (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: