Key result
In a real-world registry of patients treated with the Absorb bioresorbable vascular scaffold, the 4-year cumulative rate of target-vessel failure was 19.8%, with events continuing to accrue beyond 2 years.
Why the study?
Higher rates of device thrombosis have been reported up to 3 years after implantation of the Absorb bioresorbable vascular scaffold, prompting the evaluation of long-term clinical outcomes.
Cohort (n=135)
No
In real-world practice, target-vessel failure and scaffold thrombosis continue to accrue beyond 2 years after Absorb bioresorbable vascular scaffold implantation.
Highlights persistent long-term risks with Absorb BVS; leaves open its comparative role versus metallic stents in routine practice.
BACKGROUND: Bioresorbable scaffolds have been introduced to overcome the shortcomings of drug-eluting stents. Higher rates of device thrombosis, however, have been reported up to 3 years after implantation of the Absorb bioresorbable vascular scaffold (BVS). In the current article, we therefore report long-term clinical outcomes of the AMC Absorb Registry. METHODS AND RESULTS: In the AMC Absorb Registry, all patients who underwent a percutaneous coronary intervention with Absorb BVS implantation between 30 August 2012 and 5 August 2013 at the Amsterdam University Medical Centre-Academic Medical Centre were included. The composite endpoint of this analysis was target-vessel failure (TVF). The median follow-up of the study cohort of the AMC Absorb Registry was 1534 days. At the time of the cross-sectional data sweep the clinical status at 4 years was known in 124 of 135 patients (91.9%). At long-term follow-up, the composite endpoint of TVF had occurred in 27 patients. The 4‑year Kaplan-Meier estimate of TVF was 19.8%. At 4 years cardiac death had occurred in 4 patients (3.2%) and target-vessel myocardial infarction in 9 (6.9%) patients. Definite scaffold thrombosis occurred in 5 (3.8%) patients. We found 1 case of very late scaffold thrombosis that occurred at 911 days after device implantation in a patient who was not on dual anti-platelet therapy. CONCLUSION: In a patient population reflecting routine clinical practice, we found that cases of TVF continued to accrue beyond 2 years after Absorb BVS implantation.
No takes yet. Share an insight, caveat, or question.
Tijssen et al. (2020) conducted a cohort in Coronary artery disease requiring percutaneous coronary intervention (n=135). Absorb bioresorbable vascular scaffold (BVS) was evaluated on Target-vessel failure (composite of cardiac death, target-vessel myocardial infarction, and target-vessel revascularisation). In a real-world registry of patients treated with the Absorb bioresorbable vascular scaffold, the 4-year cumulative rate of target-vessel failure was 19.8%, with events continuing to accrue beyond 2 years.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: