Exposure to 4.7-T MRI induces translational forces in roughly a quarter of tested prostheses.
Why the study?
Are prosthetic heart valves and annuloplasty rings safe from magnetic field interactions in a 4.7 T MR system?
Are prosthetic heart valves and annuloplasty rings safe from magnetic field interactions in a 4.7 T MR system?
Most tested heart valve prostheses and annuloplasty rings are safe in 4.7 T MRI, but those containing Elgiloy (like the CE Physio Ring) may pose safety risks due to significant magnetic field interactions.
Supports compatibility of most prostheses with 4.7 T MRI; leaves open risks for Elgiloy rings requiring clinical validation.
PURPOSE: To assess the magnetic field interactions on 11 heart valve prostheses and 12 annuloplasty rings subjected to a 4.7 T MR system. MATERIALS AND METHODS: Ex vivo testing was performed to evaluate translational and rotational forces using previously described techniques. RESULTS: Seventeen out of 23 prostheses showed zero interaction with the magnetic field. Translational forces with deflection angles of 2-20 degrees were demonstrated in six prostheses. Only two heart valves and two annuloplasty rings demonstrated rotational forces. The Carpentier Edwards (CE) Physio Ring, which contains Elgiloy, demonstrated deflection angles three times greater than those previously measured at 1.5 T. Furthermore, there was a direct relationship between increasing prosthesis size and increasing translational force. All heart valve prostheses attracted to the magnetic field were slightly paramagnetic/weakly ferromagnetic. CONCLUSION: Twenty-three heart valve prostheses evaluated for MRI are considered safe in static fields up to 4.7 T based on current safety criteria. However, the CE Physio Ring appeared to develop an increasing magnetism upon re-entry into the MR system. We conclude that prostheses made from Elgiloy may not be acceptable for patients in an MR environment of > or =4.7 T. Further investigations are required to confirm the safety of Elgiloy.
No takes yet. Share an insight, caveat, or question.
Edwards et al. (2005) studied Prosthetic heart valves and annuloplasty rings (n=23). 4.7 T magnetic field was evaluated on Translational and rotational forces. Exposure to a 4.7 T magnetic field resulted in zero interaction for 17 of 23 tested prostheses, while 6 demonstrated translational forces with deflection angles of 2 to 20 degrees.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: