Key result
Right atrial leads at Bachmann's bundle are linked to ~91% less far-field R-wave sensing versus RAA.
Why the study?
Does right atrial lead implantation at the Bachmann's bundle area reduce far field R-wave sensing compared to the right atrial appendage in patients treated with multisite atrial pacing?
Cohort (n=139)
Does right atrial lead implantation at the Bachmann's bundle area reduce far field R-wave sensing compared to the right atrial appendage in patients treated with multisite atrial pacing?
Implanting the right atrial lead at the Bachmann's bundle area significantly reduces far field R-wave sensing and improves sensing safety margins compared to the traditional right atrial appendage position.
Answers are grounded in this paper and cite their sources.
BB lead placement was associated with lower FFRS incidence; leaves open whether this improves device performance in prospective trials.
AIMS: The aim of the study was to investigate far field R-wave sensing (FFRS) rate and characteristics at different right atrial (RA) positions in patients treated with multisite atrial pacing, with the RA lead implanted at the Bachmann's bundle (BB) area in 69 patients, in comparison to RA appendage (RAA) in 70 patients. METHODS AND RESULTS: All measurements were done during sinus rhythm in supine patients, with unipolar (UP) and bipolar (BP) sensing configuration. The presence, amplitude threshold (FFRS trsh) and FFRS timing were determined. Sensing safety margin was defined as the ratio of sensed P-wave vs. FFRS trsh, for both the minimal (Pmin) and the mean (Pmean) P-wave amplitude. At both atrial locations BP sensing was superior to UP in FFRS rejection (P < 0.0001). At 0.5 mV sensitivity level (BP) FFRS occurred in 1% of patients at the BB site vs. 11% at the RAA (P = 0.01). FFRS trsh (BP) was 0.2 +/- 0.1 mV at the BB vs. 0.4 +/- 0.3 mV in the RAA position (P < 0.0001). Sensing safety margin, when determined for the Pmin amplitude was > or =5 in 99% of patients from the BB group, in comparison to 66% of RAA patients (P < 0.0001), in whom it was <2 in 13%. Even with the use of BP leads equipped with a 10 mm tip-to-ring spacing FFRS incidence was lower at the BB site (P < 0.01), FFRS trsh was lower (P < 0.001), and sensing safety margin was higher vs. RAA (P = 0.002). CONCLUSION: Bachmann's bundle area features optimal conditions for signal sensing, and such atrial lead positioning may offer advantages to prevent oversensing of R-wave, thus improving functioning of standard dual chamber pacemakers, ICDs and CRT-Ds.
No takes yet. Share an insight, caveat, or question.
Lewicka et al. (2008) conducted a cohort in Multisite atrial pacing (n=139). Right atrial lead implantation at the Bachmann's bundle area vs. Right atrial lead implantation at the right atrial appendage was evaluated on Far field R-wave sensing incidence at 0.5 mV sensitivity level (bipolar) (p=0.01). Right atrial lead implantation at the Bachmann's bundle area resulted in a lower incidence of far field R-wave sensing compared to the right atrial appendage.