Key result
SoloPath expandable sheath achieves ~100% successful bail-out TAVR deployment following failed NovaFlex insertion.
Why the study?
Does the SoloPath Balloon Expandable TransFemoral Introducer enable successful TAVR deployment in patients with failed NovaFlex sheath insertion due to vessel tortuosity and calcifications?
Case Report (n=6)
Does the SoloPath Balloon Expandable TransFemoral Introducer enable successful TAVR deployment in patients with failed NovaFlex sheath insertion due to vessel tortuosity and calcifications?
The SoloPath balloon-expandable sheath can be used as an effective bail-out strategy for TAVR when standard sheath insertion fails due to severe vessel tortuosity and calcification.
Answers are grounded in this paper and cite their sources.
May offer bail-out option after sheath failure in TAVR; hypothesis-generating case report requiring prospective validation.
OBJECTIVE: Vascular complications remain the main problem of transcatheter aortic valve replacement, despite downsizing of delivery catheter diameter and judicious patient selection. In case of diffusely calcified and/or very tortuous access vessels, the SoloPath Balloon Expandable TransFemoral Introducer (Onset Medical Corporation, Irvine, CA USA), providing the smallest insertion profile of any sheath in its class, may enable exceptional trackability and offer the largest working diameter, thus helping to prevent feared vascular complications. We report the successful use of SoloPath sheath to face six cases of NovaFlex (Edwards Lifesciences Inc, Irvine, CA USA) sheath failed insertion due to vessel tortuosity and calcifications. METHODS: In six patients enrolled for transfemoral transcatheter Edwards SAPIEN XT replacement, resistance to NovaFlex sheath pushing was encountered, subsequent to the bending of the distal portion of the sheath due to vessel tortuosity and calcifications. Damaged Edwards sheaths were exchanged for a SoloPath sheath with a 21-French inner diameter. SoloPath Controlled Deployment Technology design provides the introducer, once expanded, with great radial and pillar resistance. RESULTS: The unexpanded SoloPath sheath-dilator combination tracked without resistance or kinking over a stiff guide wire through the iliofemoral axis and then was expanded. All transcatheter prostheses were successfully deployed within the native annulus, showing good hemodynamic function. CONCLUSIONS: We report six successful bail-out procedures with the SoloPath Balloon Expandable TransFemoral sheath after unsuccessful introduction of the NovaFlex device. If similar results are confirmed, use of the SoloPath during transcatheter aortic valve replacement may help prevent vascular complications and eliminate the need for other preoperative steps (eg, vascular stenting), therefore reducing procedural mortality, surgical expense, and patient recovery time.
No takes yet. Share an insight, caveat, or question.
Fusari et al. (2012) conducted a case report in Transcatheter aortic valve replacement with failed sheath insertion (n=6). SoloPath Balloon Expandable TransFemoral Introducer sheath vs. NovaFlex sheath (failed) was evaluated on Successful deployment of transcatheter prostheses within the native annulus with good hemodynamic function. The SoloPath Balloon Expandable TransFemoral sheath allowed successful bail-out deployment of transcatheter prostheses in all 6 cases following failed NovaFlex sheath insertion during TAVR.