Key result
Successful peroperative defibrillation testing predicts ~95% success on subsequent predischarge tests.
Why the study?
Does a predischarge defibrillation test provide additional value in patients with a successful peroperative test during first-time ICD implantation?
Cohort (n=30)
Does a predischarge defibrillation test provide additional value in patients with a successful peroperative test during first-time ICD implantation?
Predischarge defibrillation testing may be safely omitted in patients who have a successful peroperative test with a 10 J safety margin.
Answers are grounded in this paper and cite their sources.
Supports omitting predischarge testing after successful peroperative ICD testing; leaves open need for randomized confirmation of safety.
An internal cardioverter defibrillator (ICD) is normally extensively tested during implantation. The necessity of retesting prior to discharge of the patient is a matter of debate. In our material of 30 patients undergoing first-time implantation of a transvenous internal defibrillator system, we retrospectively compare the predischarge defibrillation test with the peroperative test. A successful peroperative defibrillation test with no failed shocks at 10 J below maximal energy level was followed by a successful predischarge test with the same safety margin in 18/19 patients, while one patient required a maximal energy ICD shock for conversion at the predischarge test. We conclude that the predischarge defibrillation test can be omitted if the peroperative test was successful, with no failed shocks at 10 J below maximal energy level and if the shock therapy is set to maximal energy level.
No takes yet. Share an insight, caveat, or question.
Abolghasem Saraj Milos Kesek (2000) conducted a cohort in First-time implantation of a transvenous internal defibrillator system (n=30). Successful peroperative defibrillation test was evaluated on Successful predischarge test with the same safety margin (no failed shocks at 10 J below maximal energy level). A successful peroperative defibrillation test with no failed shocks at 10 J below maximal energy level was followed by a successful predischarge test in 18 of 19 patients (94.7%).
Synapse has enriched 10 related papers. Consider them for comparative context: